Classical Least Squares-Assisted Mid-Infrared (MIR) Laser Spectroscopy Detection of High Explosives on Fabrics
Cargando...
Archivos
Fecha
2019
Autores
Pacheco-Londoño, Leonardo C.
Aparicio-Bolaño, Joaquín A.
Galán-Freyle, Nataly J.
Román-Ospino, Andrés D.
Ruiz-Caballero, Jose L.
Hernández-Rivera, Samuel P.
Título de la revista
ISSN de la revista
Título del volumen
Editor
Society for Applied Spectroscopy
Resumen
Mid-infrared (MIR) laser spectroscopy was used to detect the presence of residues of high explosives (HEs) on fabrics. The
discrimination of the vibrational signals of HEs from a highly MIR-absorbing substrate was achieved by a simple and fast
spectral evaluation without preparation of standards using the classical least squares (CLS) algorithm. Classical least
squares focuses on minimizing the differences between the spectral features of the actual spectra acquired using MIR
spectroscopy and the spectral features of calculated spectra modeled from linear combinations of the spectra of neat
components: HEs, fabrics, and bias. Samples in several combinations of cotton fabrics/HEs were used to validate the
methodology. Several experiments were performed focusing on binary, ternary, and quaternary mixtures of TNT, RDX,
PETN, and fabrics. The parameters obtained from linear combinations of the calculated spectra were used to perform
discrimination analyses and to determine the sensitivity and selectivity of HEs with respect to the substrates and to each
other. However, discrimination analysis was not necessary to achieve successful detection of HEs on cotton fabric substrates.
The RDX signals (mRDX>0.02 mg) on cotton were used to calculate the limit of detection (LOD). The signalto-
noise ratios (S/N) calculated from the spectra of cotton dosed with decreasing masses of RDX until S/N&3 resulted in
a LOD of 15–33 mg, depending on the vibrational band used. Linear fits generated by comparing the mass dosed RDX with
the fraction predicted were also used to calculate the LOD based on the uncertainty of the blank and the slope. This
procedure resulted in a LOD of 58 mg. Probably the most representative value of the method LOD was calculated using an
interpolation of a threshold determined using the predicted average value for the blank plus 3.28 times the standard
deviations (p-value threshold) for low surface dosages of RDX (LOD¼40 mg). The contribution demonstrates that to
achieve HE detection on fabrics using the proposed algorithm, i.e., determining the presence/absence of HEs on the
substrates, the library must contain the spectra of HEs, substrates, and potential interferents or that these spectra be
added to the models in the field. If the model does not contain the spectra of the fabric components, there is a high
probability of finding false positives for clean samples (no HEs) and a low probability for failed detection in samples with
HEs. More work will be required to demonstrate that these new approaches to HE detection work on real-world samples
and when contaminating materials are present in the samples.
Descripción
Palabras clave
Quantum cascade laser spectroscopy, QCL, High explosives, HEs, Classical least squares, CLS, Natural and synthetic fabrics, Discriminant analysis, DA